Fluke PM-3370B Manual do Utilizador Página 71

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 297
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 70
3 - 52 USING THE COMBISCOPE INSTRUMENTS
dBm - 50 offset calculation:
From the Vrms offset value the dBm-50 offset value is calculated as follows:
Example for attenuator setting 0.5 V/div.:
dBm - 600 offset calculation:
From the Vrms offset value the dBm-600 offset value is calculated as follows:
Example for attenuator setting 0.5 V/div.:
dBµV offset calculation:
From the Vrms offset value the dBµV offset value is calculated as follows:
Example for attenuator setting 0.5 V/div.:
dBm 50 offset 20 * Vrms offset 0,2236068()log
10
=
P *R() 1E-3 *50()0,2236068==
Note:
dBm 50 offset 20 * 1,118034 0,2236068()log
10
13,9794==
dBm 600 offset 20 * Vrms offset 0,7745967()log
10
=
P *R() 1E-3 *600()0,7745967==
Note:
dBm 600 offset 20 * 1,118034 0,7745967()log
10
3,1875874==
dBµV offset 20 * Vrms offset 1.0E-6()log
10
=
Note: 0 dBµV = 1 µV (1.0E-6 V) at 50 impedance.
dBµV offset 20 * 1,118034 1E-6()log
10
120,9691==
Vista de página 70
1 2 ... 66 67 68 69 70 71 72 73 74 75 76 ... 296 297

Comentários a estes Manuais

Sem comentários