COMMAND REFERENCE 4 - 97
SYSTem:COMMunicate:SERial:CONTrol:DTR
SYSTem:COMMunicate:SERial:CONTrol:RTS
Syntax:
SYSTem:COMMunicate:SERial:CONTrol:DTR ON | STANdard
SYSTem:COMMunicate:SERial:CONTrol:RTS ON | STANdard
ON Selects the "3 wire" option. The DTR or RTS line is
always asserted.
STANdard Selects the "7 wire" option.
Query form:
SYSTem:COMMunicate:SERial:CONTrol:DTR?
SYSTem:COMMunicate:SERial:CONTrol:RTS?
Response:
ON | STAN
ON "3 wire" DTR/RTS control.
STAN "7 wire" DTR/RTS control.
Description:
Controls the DTR (Data Terminal Ready) or RTS (Request To Send) line of the
EIA-232-D (RS-232-C) interface. This command has the same effect as selecting
"3 wire" or "7 wire" via front panel control. The RTS (Request To Send) line control
is coupled to the DTR (Data Terminal Ready) line control.
After a
*
RST command, the DTR/RTS control remains unchanged.
After power on, the oscilloscope is in its local state (controlled via front panel).
Example:
Send
SYSTem:COMMunicate:SERial:CONTrol:DTR ON Selects the "3
wire" control.
Front panel compliance:
The SYSTem:COMMunicate:SERial:CONTrol command is the remote equivalent
of the front panel REMOTE SETUP - RS232 SETUP option of the UTILITY menu.
Figure 4.1 Local/remote control