COMMAND REFERENCE 4 - 23
*
RST Reset
Syntax:
*
RST
Description:
The
*
RST command resets the instrument. The hardware and software of the
instrument is initialized without affecting any of the IEEE interface conditions. The
instrument turns into a fixed setup, which is optimized for remote operation. This
fixed setup is different from the setup that can be recalled via the front panel
softkeys and the SETUPS menu, which is optimized for local control.
The
*
RST command affects the following:
•
Sets the following instrument settings, independent of the past history:
Sweep time 10 ms (total acquisition)
Ω
Polarity NORMal (INV OFF)
Video signal polarity POSitive
625 video lines per frame
Low-pass cutoff frequency 0 Hz (DC coupling)
High-pass cutoff frequency bandwidth (100/200 MHz)