701/702
Users Manual
56
and connect the calibrator to the circuit to be
tested. Figure 20 shows an example.
2. If you want to automatically stop ramping if a trip
condition is detected, connect a voltage trip
circuit to the V MEAS jacks or a continuity trip
circuit to the mA
RTD MEAS jacks. (Continuity
detection is not available when sourcing current.)
3. If necessary, press M for SOURCE mode.
4. Set the calibrator for the desired source value as
previously described.
5. If you want to ramp the output in % of scale, set
% of scale as described previously under
“Sourcing in Percent of Scale.”
6.
Press the More Choices softkey.
7.
Press the Ramp softkey. The display changes to
the following: