COMMAND REFERENCE 4 - 123
Example:
Send
CONFigure:AC (@2) Configures AC-RMS CH2.
Send
SENSe:SWEep:REALtime ON Sets real-time mode on.
Send
TRIGger:SOURce INTernal2 Trigger source becomes
channel 2.
Send
TRIGger:LEVel .02 Trigger level becomes 20 mV.
Send
TRIGger:SLOPe EITHer Triggering is done at positive
(rising) and negative (falling)
trigger edges.
Send
INITiate Initiates acquisition.
Send
FETch:AC? (@2) Fetches AC-RMS value.
Read
<AC-RMS voltage> Reads AC-RMS value.
Front panel compliance:
The TRIGger:SLOPe command is the remote equivalent of the front panel TRIG1,
TRIG2, TRIG3, and TRIG4 keys and the TRIGGER MAIN TB edge option of the
TRIGGER menu.