5020A
Users Manual
6-36
The following parameters are set when this command is issued.
Sets the measurement period to 2 seconds.
Sets the measurement averaging on.
Sets printing to the serial port
Sets the temperature units to C.
Sets the temperature resolution to 2 decimal places and the relative humidity
resolution to 1 decimal place.
Stops auto logging.
Sets the statistical functions.
SYSTem:BOOT:VERSion?
Returns the boot ROM version.
Example: SYST:BOOT:VERS?
Response: 0.31
SYSTem:ERRor?
Returns a message from the system error queue.
Example: SYST:ERR?
Response: 0,"No error"
The error queue may contain up to ten messages. The messages are reported in the order
they occurred. Reading a message from the queue removes the message so the next
message can be read. If more than ten errors occur before being read, the tenth error in
the queue is -350,"Queue overflow" and all but the first nine errors are discarded. All
errors cause the Error Bit (ERR, bit 2) for the Status BYTE Register to be set. Refer to
Troubleshooting for the error messages.
SYSTem:CODE:VERSion?
Returns the firmware version.
Example: SYST:CODE:VERS?
Response: 1.00
SYSTem:VERSion?
Returns the SCPI version number.
Example: SYST:VERS?
Response: 1994.0
UNIT:TEMPerature?
Returns the system temperature unit.
Example: UNIT:TEMP?
Response: C
The response is either C for Celsius or F for Fahrenheit.